2501-1 and 2501-4 Remote Airborne Particle Counter
Particles Plus Remote Particle Counters
9000 Series vs. 2000 Series Comparison
Models 2501-1 and 2501-4
0.5 – 5 µm @ 0.1 CFM
The Particles Plus® 2501-1 Remote Particle Counter measures 0.5 μm and 5.0 μm with a flow rate of 0.1 CFM (2.83 LPM). Integration is easy into a real-time monitoring or building management system via Modbus RTU or ASCII. The 2000 Series is the smallest remote particle counters on the market and has easy mounting options. The Particles Plus® 2501-1 units comply with ISO 21501-4 and include a 1 year limited warranty.
The Particles Plus® 2501-4 Remote Particle Counter measures 0.5 μm, 0.7 μm, 1.0 μm & 5.0 μm with a flow rate of 0.1 CFM (2.83 LPM). Integration is easy into a real-time monitoring or building management system via Modbus RTU or ASCII. The 2000 Series is the smallest remote particle counters on the market and has easy mounting options. The Particles Plus® 2501-4 units comply with ISO 21501-4 and include a 1 year limited warranty.
Download Data sheets:
Measures 0.5 µm and 5 µm |
0.1 CFM (2.83 LPM) flow rate with onboard pressure loss alarm |
Long life laser diode technology |
World's smallest 0.1 CFM remote particle counter |
Measures 2 channels of simultaneous data |
MODBUS™ RTU or ASCII Outputs |
Connect using RS485 or RS232 |
Concentration limit of 500,000 per ft³ |
LED Status Indication |
Complies with ISO 21501-4 and JIS B9921 standards |
Seamless integration into a facility monitoring system |
Light weight stainless steel enclosure |
1 year limited warranty |
Channel Sizes | 0.5 μm and 25.0 μm |
Counting Efficiency | 50% at 0.5 μm; 100% for particles > 0.75 μm (per JIS) |
Concentration Limits | 500,000 particles / ft3 at 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count Level | <1 count / 5 minutes (per JIS B9921) |
Flow Rate | 0.1 CFM (2.83 LPM) |
Flow Control | Critical orifice |
Vacuum Requirements | External vacuum >15" (38.1 cm) of Hg Calibration NIST Traceable |
Sample Probe/Tubing | Isokinetic sampling probe |
Communication Interface | Modbus™ RTU or ASCII outputs (specified upon order) |
Power | 9-30 VDC |
Dimensions (L x W x H) | 1.8” x 3.7“ x 1.5” (4.5 cm x 9.3 cm x 3.8 cm) does not include probe |
Weight | 0.62 lb (0.28 kg) |
Standards | ISO 21501-4 and JIS B9921 |
Warranty | 1 year limited warranty |
Operating Conditions | 5° to 35°C 20% to 95%RH non-condensing |
Storage Conditions | -20° to 50°C Up to 98%RH non-condensing |
Optional Accessories | Isokinetic sampling probe, purge filter, data cable, vacuum and sample tubing |
Measures 0.5 μm, 0.7 μm, 1.0 μm & 5.0 μm |
0.1 CFM (2.83 LPM) flow rate with onboard pressure loss alarm |
Long life laser diode technology |
World's smallest 0.1 CFM remote particle counter |
Measures 2 channels of simultaneous data |
MODBUS™ RTU or ASCII Outputs |
Connect using RS485 or RS232 |
Concentration limit of 500,000 per ft³ |
LED Status Indication |
Complies with ISO 21501-4 and JIS B9921 standards |
Seamless integration into a facility monitoring system |
Light weight stainless steel enclosure |
1 year limited warranty |
Channel Sizes | 0.5 μm, 0.7 μm, 1.0 μm & 5.0 μm |
Counting Efficiency | 50% at 0.5 μm; 100% for particles > 0.75 μm (per JIS) |
Concentration Limits | 500,000 particles / ft3 at 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count Level | <1 count / 5 minutes (per JIS B9921) |
Flow Rate | 0.1 CFM (2.83 LPM) |
Flow Control | Critical orifice |
Vacuum Requirements | External vacuum >15" (38.1 cm) of Hg Calibration NIST Traceable |
Sample Probe/Tubing | Isokinetic sampling probe |
Communication Interface | Modbus™ RTU or ASCII outputs (specified upon order) |
Power | 9-30 VDC |
Dimensions (L x W x H) | 1.8” x 3.7“ x 1.5” (4.5 cm x 9.3 cm x 3.8 cm) does not include probe |
Weight | 0.62 lb (0.28 kg) |
Standards | ISO 21501-4 and JIS B9921 |
Warranty | 1 year limited warranty |
Operating Conditions | 5° to 35°C 20% to 95%RH non-condensing |
Storage Conditions | -20° to 50°C Up to 98%RH non-condensing |
Optional Accessories | Isokinetic sampling probe, purge filter, data cable, vacuum and sample tubing |