The 9300P-OEM measures up to 30 channels of particle count data from 0.3 to 25.0 µm with a flow rate of 1.2 LPM.The assembly is mounted on a plate for ease of integration. The unit includes a particle sensor, particle counter board, integrated vacuum pump and manifold with pre/post filters, real-time clock/calendar with battery backup, logging and data storage for up to 45,000 sample records, communicates via UART and operates from 7-15 vdc.

All Particles Plus counters meet ISO 21501-4 and JIS B9921. The 9301P-OEM ensures compliance with an on-board pulse height analyzer.

9301P-OEM: Measures 0.3 µm to 25 µm
1.2 LPM flow rate
Long life laser diode technology
Measures up to 30 channels of simultaneous data
Particle concentrations up to 30,000,000 Particles/ft3 @ I 0% coincidence loss
Internal vacuum pump with manifold and pre/post filters
Stores up to 45,000 sample records for on-board data redundancy
User-selectable channel sizes
Complies with ISO 21501-4 and JIS B9921 standards
Assembly mounted on back plate for ease of integration
I year warranty. Extended warranties available.
Model9300P-OEM
Size Range0.3 to25 µm
Size ChannelsFactory calibrated at 0.3, 0.5, 0. 7, 1.0, 2.5, and 5.0 µm
Flow Rate / Accuracy1.2 LPM / +/-3%
Concentration Limit30,000,000 Particles/ft3 @ 10% coincidence loss
Aerosol Concentration Range0.01 to 20,000 µg/m3
Light SourceLong life laser diode
Resolution0.00003 µg/m3
Zero StabilityNo appreciable drift
CalibrationN 1ST traceable
Internal Vacuum PumpInternal pump with automatic flow control
Number of Channels6to30
Custom Size ChannelsCalibration for custom size channels available
AirflowInternally monitored
AlarmAdjustable built-in alarm
Communication ModesModbus (RTU/ASCII) via 3.3v UART
StandardsISO 21501-4 and JIS B9921
Instrument CalibrationRecommended minimum once every three years
Dimensions (L x W x H)3.41" x 4.89” x 1.57” (8.7 cm x 12.4 cm x 4.0 cm) includes inlet
Weight13 oz
Optional AccessoriesStandard exhaust filter, high efficiency exhaust filter
Data Storage65,000 sample records (rotating buffer) with environmental data
Sample Time1 second to 99 hours
Power7 - 15 Vdc (< 2 watts max)
Operating Conditions41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing RH
Storage Conditions32° - 122°F (0° - 50°C) / < 98% non-condensing RH
Warranty1 year limited warranty. Extended warranties available.