2501 Remote Airborne Particle Counter
Particles Plus Remote Particle Counters
Model 2501
0.5 – 25.0 µm @ 0.1 CFM
The 2501 particle counter provides a range of particle size measurement from 0.5 to 25.0 μm at a flow rate of 0.1 CFM (2.83 LPM). These sensors allow for efficient particle counting with specifications that exceed the ISO 21501-4 and JIS B9921 calibration standards.
These sensors have 4 fixed particle size channels and are equipped with an integral critical orifice and flow sensor to provide a stable flow rate from an external vacuum source. Integration into a building automation or facility monitoring system is easy via isolated RS-485 Modbus communication.
2301: Measures 0.3 μm to 25.0 μm |
2501: Measures 0.5 μm to 25.0 μm |
0.1 CFM (2.83 LPM) flow rate |
Long life laser diode technology |
Measures up to 4 channels of simultaneous data |
Particle concentrations up to 15,000,000 Particles/ft3 @ 10% coincidence loss |
Internal critical orifice |
Connect using Modbus RTU/ASCII over isolated RS-485 |
Complies with ISO 21501-4 and JIS B9921 standards |
Easy to clean and wipe down with minimal particle traps |
Versatile mounting options |
User configurable alarm light |
LED Indicators |
Seamless integration into a facility monitoring system |
Lightweight stainless steel enclosure |
2 year limited warranty. Extended warranties available. |
Model | 2501 |
---|---|
Size Range | 0.5 to 25.0 μm |
Size Channels | Factory calibrated at 0.5, 1.0, 2.5, 5.0 μm |
Number of Channels | 4 |
Counting Efficiency | 50% @ 0.5 μm; 100% for particles > 0.75 μm per JIS |
Flow Rate | 0.1 CFM (2.83 LPM) |
Concentration Limits | 15,000,000 Particles/ft3 @ 10% coincidence loss |
Light Source | Long life laser diode |
Zero Count | < 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. |
Calibration | NIST traceable |
Vacuum Requirements | External vacuum > 15" (38.1cm) of Hg |
Airflow | Internally monitored |
Alarm | User configurable alarm light |
Communication Modes | MODBUS™ RTU or ASCII outputs (over isolated RS-485) |
Standards | ISO 21501-4 and JIS B9921 |
Instrument Calibration | Recommended minimum once per year |
External Surface | Stainless steel |
Dimensions (L x W x H) | 3.59” x 1.83” x 5.01” (9.1 cm x 4.7 cm x 12.7 cm) including probes and connectors |
Weight | 1.65 lb. (0.75 kg) |
Accessories | Wiring Connector, operating manual on USB flash drive |
Optional Accessories | Power Supply, Remote Mounting Bracket, Isokinetic Probe, Filter Assembly, and Sample & Vacuum Tubing |
Sample Time | 1 second to 18 hours |
Power | 9 - 24 VDC (< 2.5 watts) |
Operating Conditions | 41° to 104°F (5° to 40°C) / 20% to 95% non-condensing |
Storage Conditions | 32° to 122°F (0° to 50°C) / Up to 98% non-condensing |
Warranty | 2 year limited warranty. Extended warranties available. |
Part Number: AS-99081 | 0.1 CFM (2.83 LPM) Part Number: AS-99002A | 100-240 VAC 50/60Hz Part Number: EE-80148 | LPM) Stainless Steel Part Number: PS-12022 |
Bev-A-Line XX per foot Part Number: AS-99036 | Urethane per foot Part Number: AS-99018 | Line XX per foot Part Number: AS-99037 | Part Number: AS-99019 |
Part Number: AS-99088A | Series with Pump User Manual (Printed) Part Number: MN-24028P | Series Validation Manual Part Number: MN-24034 | User Manual and PPD-9K Software (USB Key) Part Number: MN-24028 |