9510 Remote Airborne Particle Counter
Particles Plus Remote Particle Counters
9000 Series vs. 2000 Series Comparison
9510 Modbus
0.5 – 25.0 µm @ 1.0 CFM
The 9510 Modbus particle counters provide a range of particle size measurement from 0.5 to 25.0 μm at a flow rate of 1.0 CFM (28.3 LPM). These sensors allow for efficient particle counting with specifications that exceed the ISO 21501-4 and JIS B9921 calibration standards.
These sensors have 6 user-selectable particle size channels (up to 30 channels optional) and are equipped with an integral critical orifice and flow sensor to provide a stable flow rate from an external vacuum source. Integration into a building automation or facility monitoring system is easy via isolated RS-485 Modbus communication.
Measures 0.5 µm to 25 µm |
1.0 CFM (28.3 LPM) flow rate |
Long life laser diode technology |
Measures up to 30 channels of simultaneous data |
Particle concentrations up to 1,500,000 Particles/ft3 @ 10% coincidence loss |
Internal critical orifice |
User-selectable channel sizes |
Stores up to 65,000 sample records for on-board data redundancy |
(Optional) temperature and relative humidity probe available |
Connect using Modbus RTU/ASCII over isolated RS-485 |
Complies with ISO 21501-4 and JIS B9921 standards |
Easy to clean and wipe down with minimal particle traps |
Versatile mounting options |
Alarm light |
LED Indicators |
Seamless integration into a facility monitoring system |
Lightweight stainless steel enclosure |
2 year limited warranty. Extended warranties available. |
Size Range | 0.5 µm to 25 µm |
Size Channels | Factory calibrated at 0.5, 1.0, 2.5, 5.0, 10.0 µm |
Number of Channels | 6 channels (up to 30 optional) |
Counting Efficiency | 50% @ 0.5µm; 100% for particles > 0.75 µm per JIS |
Flow Rate | 1.0 CFM (28.3 LPM) |
Concentration Limits | 1,500,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 3,000,000 particles/ft3 @ 10% coincidence (as tested and validated) |
Light Source | Long life laser diode |
Zero Count | < 1 count / 5 minutes (< 1 particles / 15ft3). No fault count subtraction. |
Alarms | Channel alarms on Raw counts, concentrations or mass (alarms on environmental sensors optional) |
Calibration | NIST traceable |
Vacuum Requirements | External vacuum > 15” (38.1cm) of Hg |
Airflow | Internally monitored |
Configuration/Download | USB mini-B |
Alarm | Alarm LED ring |
Communication Modes | MODBUS™ RTU or ASCII outputs (over isolated RS-485) |
Standards | ISO 21501-4 and JIS B9921 |
Instrument Calibration | Recommended minimum once per year |
External Surface | Stainless steel |
Dimensions (L x W x H) | 3.59” x 1.83” x 5.01” (9.1 cm x 4.6 cm x 12.7 cm) including probes and connectors |
Weight | 1.08 lb. (494 grams) |
Accessories | Wiring Connector, operating manual on USB flash drive |
Optional Accessories | Power Supply, Remote Mounting Bracket, Stainless Steel Barbed Isoprobe (1.0 CFM/28.3 LPM), Filter Assembly, and Sample & Vacuum Tubing |
Buffer Memory | 65,000 sample records (rotating buffer) including particle count data and environmental data |
Sample Time | 1 second to 99 hours |
Power | 9 - 24 VDC (< 1.5 watts) |
Operating Conditions | 41° to 104°F (5° to 40°C) / 20% to 95% non-condensing |
Storage Conditions | 32° to 122°F (0° to 50°C) / Up to 98% non-condensing |
Warranty | 2 year limited warranty. Extended warranties available. |
Part Number: AS-99081 | (28.3 LPM) Part Number: AS-99041A | 100-240 VAC 50/60Hz Part Number: EE-80148 | (28.3 LPM) Stainless Steell Part Number: PS-12144 |
Part Number: AS-99019 | Line XX per foot Part Number: AS-99037 | Part Number: AS-99078 | Part Number: AS-99088A |
User Manual and PPD-9K Software (USB Key) Part Number: MN-24026 | User Manual (Printed) Part Number: MN-24026P | Manual Part Number: MN-24032 | |